Why do different Pocket Colorimeters for Chlorine give different results?

Dokument-ID

Dokument-ID TE492

Veröffentlichungsdatum

Veröffentlichungsdatum 13.06.2022
Frage
Why do different Pocket Colorimeters for Chlorine give different results?
Zusammenfassung
Pocket Colorimeter results on different models
Antwort
There are three different Chlorine ranges available on a Pocket colorimeters.

Low range (LR):
Chlorine, Free and Total, Chlorine, Free and Total, Low Range, DPD Method 8021 (free)/8167 (total), AccuVac Ampuls, DR300 and Pocket Colorimeter II_DOC316.53.01450Range, DPD Method 8021 (free)/8167 (total), AccuVac Ampuls, DR300 and Pocket Colorimeter II
Mid range (MR):
Chlorine, Free and Total, Mid-Range, DPD Method 10245 (free)/10250 (total), Powder Pillows, DR300 and Pocket Colorimeter II_DOC316.53.01451
High range (HR):
Chlorine, Free and Total, High Range, DPD Method, Powder Pillows, DR300 and Pocket Colorimeter II_DOC316.53.01449 (Using 10 mL pillows)
Chlorine, Free and Total, High Range, DPD Method 10069 (free)/10070 (total), Powder Pillows, DR300 and Pocket Colorimeter II_DOC316.53.01490 (Using 25 mL pillows)

There are three different DR300s that can measure Chlorine:
DR300 Pocket Colorimeter, Chlorine, Free + Total, with Box Product # LPV445.97.00110 - This instrument can test for LR and HR with 10 mL pillows (PCII equivalent Product # 5870000)
DR300 Pocket Colorimeter, Chlorine, Free + Total, MR, with Box Product # LPV445.97.62110 - This instrument can test for MR and HR with 25 mL pillows (PCII equivalent Product # 5870062)
DR300 Pocket Colorimeter, Chlorine & pH, with Box Product # LPV445.97.12110 - This instrument can only test for HR w/ 25 mL pillows (PCII equivalent Product # 5870012)

In almost all circumstances the reason that two Pocket Colorimeters give a different result on the same sample is because the two instruments are set to use a different ranges. The different ranges have differences in the test procedure. So if two different instruments are set to a different range, but the same procedure is being followed, then the results will be different.

Here are some additional articles for more information:
Which range/sample cells should be used on the low range/high range DR300 for chlorine?
Why do the high range Chlorine methods use a 25 mL powder pillow with a 5 mL sample volume?
 