Why are the results different between the low and high range program on the DR300 for chlorine?

Dokument-ID

Dokument-ID TE504

Veröffentlichungsdatum

Veröffentlichungsdatum 11.07.2022
Frage
Why are the results different between the low and high range program on the DR300 for chlorine?
Zusammenfassung
Ranges on the DR300 for chlorine
Antwort
The DR300 for chlorine has a different procedure for each of the ranges; make sure to follow the correct procedure for the range selected. The reagent amount, sample volume, and sample cells are different between the two methods.

The US ENG Chlorine, Free and Total, High Range, DPD Method, Powder Pillows, DR300 and Pocket Colorimeter II_DOC316.53.01449 and US ENG Chlorine, Free and Total, Low Range, DPD Method 8021 (free)/8167 (total), AccuVac Ampuls, DR300 and Pocket Colorimeter II_DOC316.53.01450 methods can be found on the website.

This information also applies to the Pocket Colorimeter II for Chlorine (Product # 5870000)
 