How can the "Bootloader X.X.XX.XX" or "meter not configured" error on an HQD meter be removed or corrected?

Dokument-ID

Dokument-ID TE567

Veröffentlichungsdatum

Veröffentlichungsdatum 28.07.2022
Frage
How can the "Bootloader X.X.XX.XX" or "meter not configured" error on an HQD meter be removed or corrected?
Zusammenfassung
HQd meter displays "bootloader X.X.XX.XX" or "meter not configured" error
Antwort
This message will appear if the software was downloaded incorrectly or there is some sort of corruption with the current version of software. Download the software update files again. After the six folders are located independently on the USB device open the “Application” folder. Change the name of the .mot file to include an additional underscore ( _ ) before the .mot. The new file name should be “HQd_xxx_.mot”on the USB device. Perform the software update again.