What is the "System Check need rescan 100%T" option on the DR1900?

Dokument-ID

Dokument-ID TE730

Veröffentlichungsdatum

Veröffentlichungsdatum 03.07.2018
Frage
What is the "System Check need rescan 100%T" option on the DR1900?
Zusammenfassung
DR1900 Rescan 100%T function
Antwort
When the Rescan 100%T (System Check need rescan 100%T) is turned on in the DR1900, the instrument zeros itself on an air measurement prior to reading when an LCK or TNTplus program is selected. In addition, an "Air reading" option appears and enables the air zero to be performed manually. When this function is turned off, the LCK and TNTplus vials use an air blank value that was determined when the DR1900 was manufactured. The air blank can be helpful in accounting for changes in optics that can happen over time and contribute to inaccuracy in the measurement. The default setting for the Rescan 100% T option is "off".