If the Series 5500 analyzer molybdate 3 reagent is used for the spectrophotometer method 8282, how is the reagent blank accounted for?

Dokument-ID

Dokument-ID TE800

Veröffentlichungsdatum

Veröffentlichungsdatum 09.09.2020
Frage
If the Series 5500 analyzer molybdate 3 reagent is used for the spectrophotometer method 8282, how is the reagent blank accounted for?
Zusammenfassung
Spectrophotometer Method 8282 Reagent blank Series 5500 reagent
Antwort
The ULR Silica program 645 (method 8282) allows for a reagent blank value to be entered into the spectrophotometer. The reagent blank is based on the reagent part number. See below for reagent blank options (as provided in the “before starting the test” section of the procedure). *Verify the accuracy of the ULR method using a 500ug/L standard solution. The Silica Reagent 3, Molydate, 2.9L used for the series 5000 Silica Analyzer can be used for method 8282 for more information review the following article: Is there an ultra low range (ULR) silica method that utilizes the Silica Analyzer series 5000 or Series 5500 analyzer reagent set?

S5000 Silica Reagent 3, Molybdate, 2.9L Product # 199503 multiply reagent blank on label by 1.09. 
Molybdate 3 Reagent, 100mL Product # 199532 use the blank value on the bottle. 
Molybdate 3 Reagent, 1L, Product # 199553 use the blank value on the bottle. 

 