How can the battery life of the TSS Portable be improved?

Dokument-ID

Dokument-ID TE926

Veröffentlichungsdatum

Veröffentlichungsdatum 01.11.2018
Frage
How can the battery life of the TSS Portable be improved?
Zusammenfassung
TSS portable battery life
Antwort
The batteries shipped with the TSS portable have a 4-hour life when fully-charged. Importantly, the batteries must be charged when the meter is switched off, otherwise they will not be fully charged even if it appears to be “fully charged.”

The batteries are 6 NiMH AA-sized batteries (Product # LZY604), so their failure would not mean a replacement of the entire meter. They are considered a consumable component, as with our other meters, and are not grounds for replacement unless they fail outright upon first use.

Hat dir diese Antwort geholfen?

Wird gesendet...
Vielen Dank für Ihr Feedback.
Beim Senden ist ein Fehler aufgetreten. Versuchen Sie es erneut.