Does Hach have a procedure for measuring carbonates and bicarbonates?

Dokument-ID

Dokument-ID TE1372

Veröffentlichungsdatum

Veröffentlichungsdatum 10.05.2022
Frage
Does Hach have a procedure for measuring carbonates and bicarbonates?
Zusammenfassung
Carbonate and Bicarbonate from alkalinity
Antwort
Carbonates and bicarbonates can be determined by performing an alkalinity test. Alkalinity of water is due primarily to the presence of bicarbonate, carbonate, and hydroxide ions. Carbonate ions (CO 3 2-) are able to react with and neutralize 2 hydrogen ions (H+) and the bicarbonate ions (HCO 3 -) are able to neutralize hydrogen ions (H+) or hydroxide ions (OH-) present in water.

Carbonate concentration can be estimated by the phenolphthalein and total alkalinity measurements and use of the "Alkalinity Relationship Table" located in the attachment below:

Alkalinity, Phenolphthalein and Total Alkalinity Method 8203, Digital Titrator_DOC316.53.01166

How do the alkalinity methods measure hydroxides, carbonates, and bicarbonates? will provide additional information on how carbonates are measured and will provide the Alkalinity relationship table mentioned above.