Is there an alternate way to use the Test Filter Set LZV537?

Dokument-ID

Dokument-ID TE1441

Veröffentlichungsdatum

Veröffentlichungsdatum 20.09.2018
Frage
Is there an alternate way to use the Test Filter Set LZV537?
Zusammenfassung
LZV537 Procedures
Antwort

The Instructions for the Test Filter Set (Product # LZV537) state to perform multiple single wavelength readings and a wavelength scan (please see Where are the instructions for the Test Filter Set LZV537?).

Alternatively, current spectrophotometers have an Optical Check function that can be used with the Test Filter Set. To find this feature, from the main menu select System Checks>Optical Checks>Verfiication set. This menu option will allow input of the lot # information for the filters, walk through the process of checking the instrument, and give a pass/fail message at the end of the readings.