Can the Troemner weights, which are ASTM Class 1 Certified, be considered equivalent to any other E1, F1, or M1 certifications?

Dokument-ID

Dokument-ID TE1638

Veröffentlichungsdatum

Veröffentlichungsdatum 14.07.2020
Frage
Can the Troemner weights, which are ASTM Class 1 Certified, be considered equivalent to any other E1, F1, or M1 certifications?
Zusammenfassung
Troemner weights and E1, F1, M1 certifications
Antwort
The Troemner Weight set, Alloy 8 metric Precision Calibration, ASTM Class, 1, 100g - 1mg (Product # 2617600) are not equivalent to E1, F1, or M1 certifications, which are higher class certifications. More information on these certifications can be found on the manufacturers site. These weights (ASTM Class 1) are equivalent to OIML Class E2 however. Sample certificates can be viewed here. Tolerances are different between the classes so they do not have equivalency.