What can be done if there are still bubbles coming into my instrument after the installation of the bubble trap head regulator assembly?

Dokument-ID

Dokument-ID TE2020

Veröffentlichungsdatum

Veröffentlichungsdatum 06.01.2022
Frage
What can be done if there are still bubbles coming into my instrument after the installation of the bubble trap head regulator assembly?
Zusammenfassung
Bubble Trap Head Regulator Assembly, Product # 4668000, information.
Antwort
The bubble trap head regulator assembly is designed to help remove air bubbles from the sample stream. If there are still bubbles coming into the system, you can try to create more surface area for bubbles to adhere to inside of the trap, by placing 3 wiffle golf balls in the bottom of the trap and then 3 regular golf balls on top of them to hold them down. This can significantly improve the effectiveness of the bubble trap to relieve micro bubbles.

 