What does it mean if the sc controller is showing "replace sensor" warning for my pHD probe and how do I clear it?

Dokument-ID

Dokument-ID TE2036

Veröffentlichungsdatum

Veröffentlichungsdatum 05.01.2022
Frage
What does it mean if the sc controller is showing "replace sensor" warning for my pHD probe and how do I clear it?
Zusammenfassung
DPD1P1 sensor warning information
Antwort
The sc controllers will display a warning to "replace sensor", however this does not mean you need to replace the whole sensor. This warning is a counter that comes up yearly reminding the customer its time to change the salt bridge and cell solution of the pH probe. The warning can be cleared by acknowledging the sensor warning in the diagnostics menu.
Menu > Sensor Setup > (Choose Sensor) > Diag/Test > Counters > Reset Sensor > Are You Sure? > Enter.

See the attached articles for more information on this warning
What is the procedure for changing the salt bridge on differential pH/ORP sensors?
Is there a video on how to change a salt bridge for the pHD sensors?
Why am I unable to clear Calibration Required or Replace Sensor warnings for pHD or ORP sensor?
What happens if I don't replace the salt bridge and inner fill solution on my differential pH/ORP sensor?