How often should the salt bridge and standard cell solution of a differential pH or ORP sensor be changed?

Dokument-ID

Dokument-ID TE2055

Veröffentlichungsdatum

Veröffentlichungsdatum 05.01.2022
Frage
How often should the salt bridge and standard cell solution of a differential pH or ORP sensor be changed?
Zusammenfassung
pH/ORP sensor maintenance information
Antwort
We recommend changing the salt bridge and standard cell solution at least one time a year in a typical drinking water application. The actual frequency may be more often and is process dependent. Routine replacement of the salt bridge and standard cell solution helps to prevent the standard cell solution from becoming contaminated, which can cause an offset in the sensor signal and increased difficulty in calibration.

See the following articles for more information on changing the sensor salt bridge and electrolyte solution
What happens if I don't replace the salt bridge and inner fill solution on my differential pH/ORP sensor?
What is the procedure for changing the salt bridge on differential pH/ORP sensors
What does it mean if the sc controller is showing "replace sensor" warning for my pHD probe?
Video for how to replace the salt bridge and electrolyte solution on a pHD sensor