Why are the readings different on my sc controller and PLC/SCADA system?

Dokument-ID

Dokument-ID TE2093

Veröffentlichungsdatum

Veröffentlichungsdatum 27.02.2020
Frage
Why are the readings different on my sc controller and PLC/SCADA system?
Zusammenfassung
sc controllers output configuration
Antwort
The scaling or span must match between the sc controller and the PLC for readings to match. For the sc200, go to System Setup > sc200 Setup > Output Setup > Activation. Make sure that the set low value and set high value match between the sc controller and the PLC/SCADA system.