Which pH probe works best in applications with high ammonia levels?

Dokument-ID

Dokument-ID TE2145

Veröffentlichungsdatum

Veröffentlichungsdatum 30.03.2018
Frage
Which pH probe works best in applications with high ammonia levels?
Zusammenfassung
Strong ammonia levels and how they will effect a pH sensor.
Antwort
Hach’s pHD differential electrode is nearly ideal for samples containing strong ammonia. Normal conventional reference electrodes are destroyed in strong ammonia because the silver chloride wire inside the reference will be dissolved by the ammonia. This will cause the potential to drift and destroy the electrode, leading to bad pH readings. This damage is irreversible. Hach’s pHD differential electrode is protected against this type of damage because the silver chloride wire is encased by a glass electrode that protects it from the ammonia. With the pHD, the ammonia never gets to the wire, and the probe works fine.