My CL10 chlorine sensor is reading low. What should I do?

Dokument-ID

Dokument-ID TE2197

Veröffentlichungsdatum

Veröffentlichungsdatum 06.01.2022
Frage
My CL10 chlorine sensor is reading low. What should I do?
Zusammenfassung
CL10 chlorine sensor reading low.
Antwort
If your CL10 chlorine sensor is reading low, please see the following suggestions:
1. Make sure the probe had time to condition in chlorine-containing water if it is a new sensor, or if the membrane cap was just replaced. The sensor needs 12 to 24 hours to condition before a calibration can be performed. This is to ensure the probe has stabilized to your process before adjusting the sensor
2. Make sure the probe has been calibrated and that a 2 point calibration was performed. For applications typically below 0.5 ppm a chemical zero is recommended with a process concentration calibration. For applications typically above 0.5 ppm an electrical zero is recommended with a process concentration calibration.
3. Make sure the maintenance on the probe is up to date and the cap is very secure on the end of the probe. Depending on the application, the electrolyte will need to be changed more often than others (lower concentration typically means changing the electrolyte more often)

Hat dir diese Antwort geholfen?

Wird gesendet...
Vielen Dank für Ihr Feedback.
Beim Senden ist ein Fehler aufgetreten. Versuchen Sie es erneut.