What is the difference between salt bridge kits, Product # SB-P1SV and Product # SB-P1SP?

Dokument-ID

Dokument-ID TE2282

Veröffentlichungsdatum

Veröffentlichungsdatum 05.01.2022
Frage
What is the difference between salt bridge kits, Product # SB-P1SV and Product # SB-P1SP?
Zusammenfassung
Differential pH and salt bridge Product #'s.
Antwort
SB-P1SV is the standard salt bridge used on any pHD or pHDsc sensors.  Some applications may expose these sensor's wetted materials to potential failure due to chemical incompatibility.  The standard O-rings included with the SB-P1SV are Viton material.  When the Viton material is not chemically compatible with the chemicals in an application, the SB-PF1S using perfluroelastomer O-rings is available.

Product # SB-P1SV - Salt Bridge for pHD sc and pHD, PEEK Body and Kynar (PDF) Outer Junction for PEEK Sensor with Viton O-rings

Product # SB-PF1S - Salt Bridge for pHD sc and pHD, PEEK Body and Kynar (PDF) Outer Junction for PEEK Sensor with perfluroelastomer O-rings