How do you troubleshoot a 2200PCX Particle Counter which will not communicate with a PC via OPC Datalogger sofware?

Dokument-ID

Dokument-ID TE2289

Veröffentlichungsdatum

Veröffentlichungsdatum 05.03.2018
Frage
How do you troubleshoot a 2200PCX Particle Counter which will not communicate with a PC via OPC Datalogger sofware?
Zusammenfassung
2200PCX and no communication with PC via OPC Datalogger
Antwort
Software
Use PCX Explorer software: (ProcessLink and ModIO Explorer will not work with the 2200PCX).

Hardware
For RS485 Output: Refer to RS485 wiring information in Section 3.2.5 in Operator Manual

For RS232 Output: Use J2 connector. If using an RS232-to-USB adapter for the PC:Install the driver for the adapter, provided by the manufacturer of the adapter. Locate the Com Port number assigned to the adapter by clicking “Windows Start Menu> Control Panel> System> Hardware Tab> Device Manger> Ports.” Expand the Ports branch to list available Com Ports on the PC. Com port should be set for 9600 baud, 8 data bits, no parity, 1 stop bit, no flow control.
 

Hat dir diese Antwort geholfen?

Wird gesendet...
Vielen Dank für Ihr Feedback.
Beim Senden ist ein Fehler aufgetreten. Versuchen Sie es erneut.