Why is the BODTrak™II displaying time in minutes rather than days?

Dokument-ID

Dokument-ID TE3254

Veröffentlichungsdatum

Veröffentlichungsdatum 06.09.2019
Frage
Why is the BODTrak™II displaying time in minutes rather than days?
Zusammenfassung
BODTrak™II shows minutes not days
Antwort
The time should be displayed in days; if it is not, follow these steps to restore the correct time display:
  1. From the "Instrument Setup" screen push "OFF" and the channel display screen, which displays the graph, will appear.
  2. Push "ON" and the "Range Selection" screen will appear.
  3. Push the "1" and "6" keys simultaneously.
  4. Push "OFF," then push and hold the "<" and ">" keys simultaneously to go back to the "Instrument Setup" screen to choose the proper test length.