Are the frits and guard disc on the Lachat IC column replaceable?

Dokument-ID

Dokument-ID TE4325

Veröffentlichungsdatum

Veröffentlichungsdatum 05.01.2022
Frage
Are the frits and guard disc on the Lachat IC column replaceable?
Zusammenfassung
Replacing the guard disc and frits in the column
Antwort
Yes, the frits going into the column are replaceable. The tan guard disc is Product # 28132. The green frit is Product # 28110. The guard disc is designed to be replaced when clogged and can be considered a consumable item. We do not recommend replacing the green frit in the column unless it is a last ditch effort and the column is expected to be replaced soon anyways, as removing the frit can be difficult and can cause damage to the column.

 