Can the internal sample sequencer on the 5500sc Series analyzers be configured to monitor one sample stream more often than the other streams?

Dokument-ID

Dokument-ID TE4760

Veröffentlichungsdatum

Veröffentlichungsdatum 06.01.2022
Frage
Can the internal sample sequencer on the 5500sc Series analyzers be configured to monitor one sample stream more often than the other streams?
Zusammenfassung
Configure the 5500sc Series analyzers to monitor one sample stream more often than the other streams
Antwort
Multi-stream 5500sc analyzers have stream selections for S1 through S12. Each of these points can be configured to monitor individual streams, in any preferred sequence. If any of the sequence points are not needed, they can be left blank (no stream assigned).

ADDITIONAL DETAILS:
The sequencer is configured in the SETUP SYSTEM>CONFIGURE SEQUENCER menu.
 