Why are the UVAS readings not matching Spectrophotometer readings? Expectation is for the UVAS to be within 5% of lab readings.

Dokument-ID

Dokument-ID TE4836

Veröffentlichungsdatum

Veröffentlichungsdatum 05.01.2022
Frage
Why are the UVAS readings not matching Spectrophotometer readings? Expectation is for the UVAS to be within 5% of lab readings.
Zusammenfassung
UVAS readings not matching Spectrophotometer readings
Antwort

Be sure they are all using the same method - which means not filtering the sample on the portables, and making sure the Reference is set to 'off' in the UVAS program.  Also, you have to evaluate the cleanliness of everything - glassware, sample cells, and lenses. It may be that the UVAS needs a more frequent wiping interval.

The last step would be to evaluate both systems against a known standard if necessary. Our COD standard is potassium acid phthalate, which could be diluted to make a %UVT standard. If all instruments check out fine on the standard, then we know it is a sample issue, and perhaps we should run spikes/standard additions.

 

Hat dir diese Antwort geholfen?

Wird gesendet...
Vielen Dank für Ihr Feedback.
Beim Senden ist ein Fehler aufgetreten. Versuchen Sie es erneut.