What does “open” mean when displayed on a Lachat heater controller?

Dokument-ID

Dokument-ID TE5221

Veröffentlichungsdatum

Veröffentlichungsdatum 07.03.2019
Frage
What does “open” mean when displayed on a Lachat heater controller?
Zusammenfassung
"Open" message on heater controller
Antwort
Open will be displayed if the thermocouple is not plugged in. The plug for the Type K Sheathed Thermocouple (Product # 85268) is yellow on the Lachat heater module assembly for the QC8500. If this is properly plugged in and another channel with heater is available, the heater block with attached thermocouple can be switched to see if the problem moves with the thermocouple. If the problem does not move with the thermocouple the issue is likely with the heater controller. For issues with the heater controller, the heater module assembly should be repaired or replaced.  
 