How is the ORP sensor calibrated on the MP-6 and MP-6P portable meters?

Dokument-ID

Dokument-ID TE6269

Veröffentlichungsdatum

Veröffentlichungsdatum 20.03.2018
Frage
How is the ORP sensor calibrated on the MP-6 and MP-6P portable meters?
Zusammenfassung
Calibration of ORP on MP meters
Antwort

The ORP electrodes rarely give false readings unless there is a problem in the reference electrode. For this reason, and because the calibration solutions for ORP are highly reactive and potentially hazardous, the MP meter has an electronic ORP calibration. This causes the zero point on the reference electrode to be set whenever the pH 7 calibration is done.
 
An ORP sensor uses a small platinum surface to accumulate charge without reacting chemically. That charge is measured relative to the solution, so the solution "ground" voltage comes from the reference junction. The same reference is used for both the pH and the ORP sensors. Both pH and ORP indicates 0 for a neutral solution. Calibration at zero corrects the error in the reference junction.
 
A zero calibration solution for ORP is not practical, so the MP series meters uses the offset value determined during calibrationto 7 in pH calibration (pH 7 =0 mV). Sensitivity of the ORP surface is fixed so there is no gain adjustment.

Hat dir diese Antwort geholfen?

Wird gesendet...
Vielen Dank für Ihr Feedback.
Beim Senden ist ein Fehler aufgetreten. Versuchen Sie es erneut.