The sample probe installed on the Lachat ASX autosampler appears loose. What can be done about this?

Dokument-ID

Dokument-ID TE6325

Veröffentlichungsdatum

Veröffentlichungsdatum 26.03.2019
Frage
The sample probe installed on the Lachat ASX autosampler appears loose. What can be done about this?
Zusammenfassung
Lachat sampler single probe guide plate.
Antwort
For Lachat ASX autosamplers that include a sample probe guide plate, the plate has two holes in it. A screw underneath the probe assembly can be loosened to turn the guide plate 180 degrees. This will put the other hole into position. The larger hole is needed to accommodate the dual probe used with an auto dilutor. The smaller hole is for the smaller diameter single probe.

This sampler probe guide plate is visible as item 9 below.



 