After adjusting the peak expectation window on my Lachat analyzer the peaks are not integrated properly. What is happening?

Dokument-ID

Dokument-ID TE6457

Veröffentlichungsdatum

Veröffentlichungsdatum 19.07.2018
Frage
After adjusting the peak expectation window on my Lachat analyzer the peaks are not integrated properly. What is happening?
Zusammenfassung
Lachat Peak Integration adjustment
Antwort
Verify the Threshold setting is set to zero.  This is visible in the Run Properties window under the Analytes tab.  Make sure to select the analyte your are working with from the left hand side of the analytes tab.  Threshold is the last parameter in the list.  If this is not set to zero, change to zero and rerun peak detection.  