Peaks are not being integrated properly by the Lachat software. Is there an Omnion software setting that can be adjusted to resolve this?

Dokument-ID

Dokument-ID TE6458

Veröffentlichungsdatum

Veröffentlichungsdatum 1553627215000
Frage
Peaks are not being integrated properly by the Lachat software. Is there an Omnion software setting that can be adjusted to resolve this?
Zusammenfassung
Lachat Peak Integration
Antwort
Yes. The integration can be adjusted directly in the channel data display (window in which the peaks appear).  The entire box and/or either edge of the box can be adjusted.

1.  Right-click within the channel data display, then choose Adjust Peak Expectation Window



2.  Select one of the black boxes and it will become red showing it is the box to be adjusted. 
3.  Placing the cursor within the box will show a four headed arrow. Click and then hold the left mouse button while this is displayed will allow one to move the entire box.  



4. Placing the cursor on either edge of the box will show a two headed arrow.  Click and hold the left mouse button while this is displayed will allow one to move the selected edge.  



5.  Right-click within the channel data display, once the box is properly adjusted, and choose Rerun Peak Detection.  



This will adjust the integration for all the peaks within the run.