Are there general maintenance and cleaning steps to follow for my SL1000 Portable Parallel Analyzer™ (PPA)?

Dokument-ID

Dokument-ID TE1128

Veröffentlichungsdatum

Veröffentlichungsdatum 03.01.2022
Frage
Are there general maintenance and cleaning steps to follow for my SL1000 Portable Parallel Analyzer™ (PPA)?
Zusammenfassung
Cleaning and maintenance of the SL1000 Portable Parallel Analyzer™ (PPA)
Antwort
Yes, the Chemkey® slots, optical windows, barcode window, sample detector and trays can all be inspected and cleaned. Parts should be cleaned with isopropyl alcohol and cotton tipped applicators. Care should be taken to swipe the cotton tipped applicator in one direction only, not back and forth. Water and a mild detergent can be used to clean the light shield and the optics access cover. Trays and windows should be visually inspected for clouding or buildup.

Hat dir diese Antwort geholfen?

Wird gesendet...
Vielen Dank für Ihr Feedback.
Beim Senden ist ein Fehler aufgetreten. Versuchen Sie es erneut.