Does have Hach have an instrument that can measure active chlorophyll in a water sample?

Dokument-ID

Dokument-ID TE1338

Veröffentlichungsdatum

Veröffentlichungsdatum 20.12.2021
Frage
Does have Hach have an instrument that can measure active chlorophyll in a water sample?
Zusammenfassung
The BW680 Ballast Water Handheld Fluorometer measures active chlorophyll
Antwort
The BW680 measures active chlorophyll A in systems like phytoplankton or cyanobacteria. It will not give a concentration, but will display a result of RISK FAIL, RISK HIGH, or RISK LOW. RISK FAIL means active chlorophyll is present. RISK HIGH means that chlorophyll is detected and RISK LOW means that no chlorophyll was detected. Note: this will not tell you if you pass regulations for ballast water, only if you fail. Also, the BW680 cannot tell how many chlorophyll containing organisms are in the sample since they may vary in size. Assuming all organisms are 15um, the system should be able to detect 10 phytoplankton per one milliliter of sample.