What is the relationship between Ammonia (NH3) and other Nitrogen groups?

Dokument-ID

Dokument-ID TE157

Veröffentlichungsdatum

Veröffentlichungsdatum 18.08.2020
Frage
What is the relationship between Ammonia (NH3) and other Nitrogen groups?
Zusammenfassung
Relationship between nitrogen containing compounds
Antwort
Nitrogen exists in the environment in many forms and changes forms as it moves through the nitrogen cycle. A variaty of nitrogen methods will test the different forms of nitrogen.

Provided below is a summary of the relationship between different nitrogen compounds and the nitrogen method recommended for analysis.

(TIN) = (NH3) + (NO3)
(TN) = Organic Nitrogen (All) + NH3 + NO3 + NO2
(TKN) = Organic Nitrogen (trinegative oxidative*) + NH3

*Trinegative oxidative: Will not account for azide, azine, azo, hydrazone, nitrile, nitro, nitroso, oxime, and semi-carbazone

Comparing TKN results
TKN = (approx) TN - (NO3 + NO2)
= (approx) NH3 + Organic Nitrogen

Organic Nitrogen:
TKN - NH3 = Organic Nitrogen
TN-TIN = Organic Nitrogen

Acryonym Legend:
Total Inorganic Nitrogen (TIN)
Ammonia (NH3)
Ammonium (NH4)
Nitrate (NO3)
Nitrite (NO2)
Total Kjeldahl Nitrogen (TKN)
Total Nitrogen (TN)