What happens to an LDO2 if the temperature is below 0 degrees C?
Zusammenfassung
LDO2sc Sensor and Temperature
Antwort
The temperature range is 0-50 degrees C for the LDO2 sensor. If the temperature is outside of that range the instrument is not guaranteed to function properly. When the sensor is operated outside of the temperature range, the values will be from a non-continuous realm. In other words, the numbers will be total nonsense and cannot simply be extrapolated based on readings inside the specified range. If you are at 2 ppm in 0 degrees C water, your instrument value could be anything at -1 degrees C. The readings will be nonsense and cannot be used. The temperature reading is accurate and will show from -5 to 55 degrees C, so 5 above and 5 below the range of the instrument.