How do you update the firmware of the LDO2 sc sensor?

Dokument-ID

Dokument-ID TE2296

Veröffentlichungsdatum

Veröffentlichungsdatum 05.01.2022
Frage
How do you update the firmware of the LDO2 sc sensor?
Zusammenfassung
LDO2 sc and firmware update.
Antwort
Required Hardware:
LDO2 sensors
sc200 or sc1000 controller
Software update file, available on MyDanaher
SD Card
SD Card reader for PC

Procedure:
Navigate to MyDanaher and download the software update file to the PC. From MyDanaher, click Hach> Water Quality> Common Controller Platform> Released Software> Smart Sensor Files
Power-down sc-controller.
Insert the SD Card into the SD slot on the sc-controller.
Power-up sc-controller. The sc-controller automatically creates a sub-folder, within the HACH\Firmware folder, on the SD Card for the LDO2 sensor.
Remove SD Card from controller and insert into SD Card reader. Insert SD Card reader into USB port on the PC. Copy software update file from the PC to the SD Card. Copy the file to the LDO2 sub-folder, within the HACH\Firmware folder.
Remove the SD Card from the PC and insert it into SD slot on sc-controller. On the sc-controller, press Menu> SD Card Setup> Upgrade Software> Enter.