What can cause the Astro TOC/1950+ to read zero and what can fix this?

Dokument-ID

Dokument-ID TE2311

Veröffentlichungsdatum

Veröffentlichungsdatum 21.01.2019
Frage
What can cause the Astro TOC/1950+ to read zero and what can fix this?
Zusammenfassung
Astro TOC/1950+ Reading Zero
Antwort
Zero readings can be a calibration issue. 1. Check the calibration. If the calibration has been performed when there was a leak in the system it is possible the curve is a flat line. This will cause the analyser to read zero, and this is especially true if you are able to perform a gas calibration. 2. The live value is not based on calibration, it is the raw value. Do a liquid cal to see if it will calibrate and read the standard properly. What ppm does it show? If 0, there is an issue with the bench. Recalibrate it.

Hat dir diese Antwort geholfen?

Wird gesendet...
Vielen Dank für Ihr Feedback.
Beim Senden ist ein Fehler aufgetreten. Versuchen Sie es erneut.