How often should be done the matrix corrections for the ISE online sensors?

Dokument-ID

Dokument-ID TE3433

Veröffentlichungsdatum

Veröffentlichungsdatum 06.01.2022
Frage
How often should be done the matrix corrections for the ISE online sensors?
Zusammenfassung
ISE matrix correction
Antwort
During the first month of operation a matrix correction should be performed every week. After ensuring that this interval is sufficient, it can be reduced to longer intervals.
The matrix correction should not exceed 1 month. After that period it is highly recommended to check the matrix correction.
A matrix correction should be performed whenever there is uncertainty about the quality of the readings and deviations to the lab results are detected.

.