What are the specifications for the Water Purification System Universal Filtration Cartridge?

Dokument-ID

Dokument-ID TE5381

Veröffentlichungsdatum

Veröffentlichungsdatum 05.01.2022
Frage
What are the specifications for the Water Purification System Universal Filtration Cartridge?
Zusammenfassung
Specifications for the Water Purification System Universal Filtration Cartridge.
Antwort
The specifications for the Water Purification System Universal Filtration Cartridge (Catalog number/Order code 2967100) are:
 
  • Hydrogen for strong acid cation resin mixed with weak base anion resin on a 50/50 volume basis
  • Cation resin dyed purple- changes to amber upon exhaustion
  • Removes all ions with the exception of Silica and CO2
  • Capacity 1600 grains as CaCO3
  • Effluent pH 6-7, Conductivity 10-20 µS/cm
  • Applications include:
    • Humidity/environmental chambers
    • Battery water
    • Steamers for food industry
    • Laboratories
    • Cooling loops
    • Steam irons
  • 3 1/2" Diameter x 19 3/4" Length
  • 30 psi maximum rated pressure
  • 100 °F maximum temperature
  • Shelf life of 2 years
  • Design flow rate of 7.2 GPH

Hat dir diese Antwort geholfen?