How does the 5500 sc Silica or Phosphate measure the flow rate into the analyzer?

Dokument-ID

Dokument-ID TE6008

Veröffentlichungsdatum

Veröffentlichungsdatum 06.01.2022
Frage
How does the 5500 sc Silica or Phosphate measure the flow rate into the analyzer?
Zusammenfassung
5500 sc sample flow measurement details.
Antwort
The 5500 sc analyzer uses a pressure differential technique to calculate the flow rate into the analyzer. The tubing length and characteristics of the tubing and heating coil are known and used in the calculation of the sample flow rate. When looking in the analyzer data section of the menu (menu > view data > analyzer data) pressure 1 and pressure 2 are displayed.  When the sample solenoid valve is closed these two pressures should be close to the same value.  When the sample solenoid valve opens, pressure 2 will drop to close to 0 while pressure 1 will remain the same.  The analyzer will calculate the flow rate when this pressure drop occurs.

Hat dir diese Antwort geholfen?

Wird gesendet...
Vielen Dank für Ihr Feedback.
Beim Senden ist ein Fehler aufgetreten. Versuchen Sie es erneut.