What's the difference between temperature compensation and temperature correction? (pH verses Conductivity temperature adjustments)

Dokument-ID

Dokument-ID TE6094

Veröffentlichungsdatum

Veröffentlichungsdatum 04.04.2019
Frage
What's the difference between temperature compensation and temperature correction? (pH verses Conductivity temperature adjustments)
Zusammenfassung
Temperature compensation vs correction.
Antwort
Temperature compensation is a term most common to pH measurement. This refers to using the measured temperature in the calculation of the final result instead of using a reference temperature of 25 °C. This way the result is the pH of the solution at that temperature, instead of pH of a solution at 25 °C with the measured mV.

Temperature correction is a term most common to conductivity. This refers to the measured conductivity being converted by a factor based on the measured temperature to display an estimated conductivity if the sample were to be at reference temperature (usually 25 °C) instead of the measured temperature. This can be linear (a factor of x% per °C from the reference temperature) or non-linear where the percentage factor varies depending on the difference between the measured temperature and reference temperature. Non-linear temperature correction curves are exclusively based on a standard, and can be different depending on the standard that the curve is based on. With conductivity temperature correction, there can be a different correction for the calibration than there is for measurement. Calibrations are commonly non-linear corrections, whereas the measurement corrections can be either. This is because the relationship between conductivity and temperature is more clearly defined on a standard, but can vary on samples based on sample matrix. The IntelliCAL™ CDC401 as an example will use a non-linear temperature correction curve specific to the selected standard if one of the preset standards is used for calibration. If a custom standard is used it will use and adjustable linear correction factor defaulting to 2%/°C. For measurement it uses the Non-linear NaCL standard based temperature correction curve by default, but can be changed to either non-linear natural water curve based on 442-30 natural water standards or linear with an adjustable factor.

 

Hat dir diese Antwort geholfen?

Wird gesendet...
Vielen Dank für Ihr Feedback.
Beim Senden ist ein Fehler aufgetreten. Versuchen Sie es erneut.