When using TNT817 for International Bitterness Units, the barcode on the vial does not start the program on the Hach spectrophotometer. What is causing this?

Dokument-ID

Dokument-ID TE6994

Veröffentlichungsdatum

Veröffentlichungsdatum 27.05.2017
Frage
When using TNT817 for International Bitterness Units, the barcode on the vial does not start the program on the Hach spectrophotometer. What is causing this?
Zusammenfassung
Program for TNT817 for International Bitterness Units
Antwort
The program for TNT817 for International Bitterness Units (Catalog Number/Order Code TNT817) is not associated with the barcode as it is when using other TNTplus® chemistries. When using TNT817, program 817 should be selected from the Stored Programs list. If program 817 is not found, the software on the Hach spectrophotometer should be upgraded from the Hach website.