The magnetic stirrer is not working on the AT1000 Titrator. How can this be resolved?

Dokument-ID

Dokument-ID TE7343

Veröffentlichungsdatum

Veröffentlichungsdatum 24.01.2019
Frage
The magnetic stirrer is not working on the AT1000 Titrator. How can this be resolved?
Zusammenfassung
Magnetic stirrer functions on an AT1000
Antwort
If the magnetic stirrer is not working on an AT1000, first verify that the titrator is running the current version of software by selecting Settings then Info. See also What are the instructions for performing a software update on the AT1000/KF1000?

If the instrument is running the current software version, verify that the titrator is set to use a propeller stirrer instead of the magnetic stirrer. This can be checked by going to Settings then Options, and make sure that the box for the propeller stirrer is unchecked.

Sometimes the stir bar can have enough friction on the bottom of the glass that the stirrer isn't providing enough force to get it started. Increasing the stirring speed with the up arrow or tapping the beaker can sometimes resolve the issue.

 