What can cause negative results in LCK 380 TOC?

Dokument-ID

Dokument-ID TE7457

Veröffentlichungsdatum

Veröffentlichungsdatum 02.02.2022
Frage
What can cause negative results in LCK 380 TOC?
Zusammenfassung
Measuring method in LCK 380 TOC
Antwort

As we can see in the attached manual we have to prepare for every single sample one TC and one TIC cuvette.

Both of them have to be measured as given in the manual.

The TOC will be only calculated by the subtraction of measuring result TIC from measuring result TC:               TC-TIC= TOC

If we use a  DR-instrument we have to measure both cuvettes first TC than TIC and after that we will get the final result.

(Please see and follow procedure given on page 5)

So, if we have  mixed up the cuvettes we may get some negative results.

 