At which wavelength does the DR 6000 or DR 5000 switch from using the UV lamp to the visible lamp?

Dokument-ID

Dokument-ID TE7585

Veröffentlichungsdatum

Veröffentlichungsdatum 06.01.2022
Frage
At which wavelength does the DR 6000 or DR 5000 switch from using the UV lamp to the visible lamp?
Zusammenfassung
DR 5000 DR 6000 UV Switch wavelength
Antwort
The DR 5000 and DR 6000 spectrophotometers switch from using the visible lamp at 332 nm to using the UV lamp at 331 nm by default. 
This default value can be changed by selecting "Instrument Setup" in the main menu, and editing the value in the "UV Switch:". The wavelengths that the can be selected must be between 320 and 360 nm. 

Refer to Section 5.2.9 Lamp Control in the DR 6000 Spectrophotometer User Manual  for these instructions.