What causes a "CARTRIDGE OLD" warning on the AN-ISE sc, A-ISE sc or N-ISE sc probe?

Dokument-ID

Dokument-ID TE7749

Veröffentlichungsdatum

Veröffentlichungsdatum 06.01.2022
Frage
What causes a "CARTRIDGE OLD" warning on the AN-ISE sc, A-ISE sc or N-ISE sc probe?
Zusammenfassung
"CARTRIDGE OLD" warning and ISE sc sensors
Antwort
The warning "CARTRIDGE OLD" is displayed if a new cartridge is installed and it is over a year old.

This can also happen if the wrong date and time was set up in the controller for the first Matrix Correction. When  this "old" date is more than one year in the past the message "CARTRIDGE OLD" is displayed. It is not possible to reset this warning. The only solution is to use a brand new cartridge but make sure that the time stamp from the controller is correct during the first Matrix correction.