No peaks are visible are when running samples and standards on a Lachat QuickChem 8000/8500 series instrument. What could be the cause?

Dokument-ID

Dokument-ID TE7987

Veröffentlichungsdatum

Veröffentlichungsdatum 04.03.2022
Frage
No peaks are visible are when running samples and standards on a Lachat QuickChem 8000/8500 series instrument. What could be the cause?
Zusammenfassung
No peaks observed on a Lachat QuickChem 8000/8500 series instrument.
Antwort
If no peaks are visible when testing samples and standards, try running dye as a sample. Please note that dye will not give peaks at 880 nm. If a phosphorous chemistry is being used (or any other chemistry that uses a 880 nm filter), the filter should be replaced with one from another chemistry to view the peaks.
  • If peaks are observed when running dye as a sample, the following should be considered:
    • Verify the filter used is correct for the chemistry being run.
    • Look at the reagents and standards to verify they will produce color. Try mixing the reagents and a little of the stock standard in a test tube and see if a color forms.
    • Review the manifold to see if reagent lines were mixed up either at the reagent bottles or the manifold.
    • Verify each of the reagent lines flows properly. One can get a peak with dye as a sample if the carrier flows well, regardless if the other reagent lines flow or not. However, one will not get a peak when testing actual samples and standards as all of the reagents are needed for the chemistry to work.
  • If no peaks are observed when running dye as a sample, the following should be considered:
    • Is the dye getting injected into the manifold? Check the valve timing.
    • Verify the entire manifold matches the Lachat method being followed.
    • Check the baseline. If it is at 10 V of higher, the lamp may be out, or the flow cell may be inserted incorrectly.  If these things have been verified, check to see if light is coming out of the fiber optics; you can also remove a component (filter or flow cell) from the detector while previewing to verify the baseline shifts. if it does not, there is a hardware or configuration issue.
Is there a replacement for the Lachat Dye, Universal 250ml?

Are there any suggestions or tips on troubleshooting flow problems on the Lachat QC8500 FIA instruments?

How is valve timing optimized on the Lachat QC8500 FIA?

Is there a guide to troubleshooting an elevated baseline on the Lachat Quikchem FIA instruments?

 

Hat dir diese Antwort geholfen?

Wird gesendet...
Vielen Dank für Ihr Feedback.
Beim Senden ist ein Fehler aufgetreten. Versuchen Sie es erneut.