What is the proper procedure to calibrate an In-pipe Ultrasonic sensor with the sensor height option for a Sigma 950?

Dokument-ID

Dokument-ID TE8032

Veröffentlichungsdatum

Veröffentlichungsdatum 20.12.2021
Frage
What is the proper procedure to calibrate an In-pipe Ultrasonic sensor with the sensor height option for a Sigma 950?
Zusammenfassung
Calibration of In-Pipe Ultrasonic sensor with the sensor height option on a Sigma 950.
Antwort
  1. From the Main Menu select, Option, Advanced options, Calibration, Ultrasonic Sensor.  
  2. Highlight the calibrate U-sonic, click Select.
  3. Change the transducer type to (In-Pipe) using the Change Choice button, press Accept to continue with the calibration. 
  4. Enter in the Air temperature where the sensor is located with the keypad, then press Select to continue with the calibration.
  5. Then Select sensor height option.
  6. Take a physical measurement from the sensor face to the zero-flow point in the channel. Enter the measured distance in the zero-flow distance.
  7. Press Accept when finished. The calibration is now complete  