Will the Ammonia (Nitrogen) Test Strips, 0-6.0 mg/L detect quaternary ammonium compounds or residues?

Dokument-ID

Dokument-ID TE8081

Veröffentlichungsdatum

Veröffentlichungsdatum 08.01.2019
Frage
Will the Ammonia (Nitrogen) Test Strips, 0-6.0 mg/L detect quaternary ammonium compounds or residues?
Zusammenfassung
Ammonia Test Strips and quaternary ammonium compounds and residues
Antwort
The Ammonia (Nitrogen) Test Strips, 0-6.0 mg/L (Product # 2755325) will not detect quaternary ammonium compounds or residues. The test strips work by raising the pH of the sample to above 10 which converts all ammonium ions in the sample to free (un-ionized) ammonia. Quaternary ammonium cations are permanently charged cations, independent of pH, so they do not get converted to free ammonia. Therefore, quaternary ammonium can not be detected with the ammonia test strips.
 