How do you program the 4-20 mA outputs on a Sigma 950?

Dokument-ID

Dokument-ID TE8138

Veröffentlichungsdatum

Veröffentlichungsdatum 20.12.2021
Frage
How do you program the 4-20 mA outputs on a Sigma 950?
Zusammenfassung
Programming the 4-20 mA outputs on a Sigma 950.
Antwort
  1. From the Main Menu select Options, Advanced Options, 4-20 mA.  
  2. Select the Change Choice keys to enable the 4-20 mA, then press Accept.
  3. Select the Output Channel to program (Output A or Output B) using the up and down arrow keys, then press Select.
  4. Select the Input Channel by pressing the Change Choice key, When the desired input channel is displayed, Press the Accept key.
  5. Assign a value to the 4 mA Current Value using the keypad (this value is normally 0.00 or your lowest flow number). Then press Accept.
  6. Assign a value to the 20 mA Current Value using the keypad (this value is normally the maximum flow of the channel). Then press Accept.