Does the HQd meter automatically store measurement and calibration data?

Dokument-ID

Dokument-ID TE8297

Veröffentlichungsdatum

Veröffentlichungsdatum 19.09.2018
Frage
Does the HQd meter automatically store measurement and calibration data?
Zusammenfassung
HQd meter automatic data storage
Antwort
The following types of data are stored in the data log:
  • Sample measurements: stored automatically each time a sample is measured in the Press to Read or Interval Mode. When the continuous measurement mode is used, data is stored only when Store is selected.
  • Calibrations: stored only when Store is selected at the end of a calibration. Calibration data is also stored in the IntelliCAL (R) probe. 
  • Check standard measurements: stored automatically each time a check standard is measured (in the Press to Read or Interval Mode).
When the data log becomes full (500 data points), the oldest data point is deleted when a new data point is added.

Refer to the HQd Meter Manualsection for more information regarding data management:
Where can I find the User Manual for the HQd Portable meters?
Where can I find the User Manual for the HQd Benchtop Meters?