Frage
How does the TOC QBD1200 analyzer option "Reject Outliers" work?
Zusammenfassung
QBD1200 "Reject Outliers" option
Antwort

The "Reject Outliers" option allows the user to have replicates automatically rejected (i.e. not included in reporting and statistics) based on a target Std Dev/RSD. Pretend the user has chosen to do 4 replicates for a drinking water sample and "reject outliers" with a target deviation of 3.0. In that situation, the analyzer will measure all 4 reps and check the RSD to see if the 3.0 % criteria is met. If it isn't, the analyzer software then determines which group of three replicates results in the best RSD and rejects the fourth. If the user chose 5 reps, the analyzer would do the same thing but only reject enough replicates to meet the target RSD %. The analyzer will never discard below 3 reps, though. That's why the option is disabled if the user selects 3 or fewer reps. Below are some examples of what this looks like if the user selects a 3.0% target.

 

One other thing to note is that the target becomes standard deviation in ppb if the sample average is below 100 ppb. That fits with our reported reproducibility stat of 3.0 ppb std dev or 3.0% RSD and the reasoning behind that change below 100 ppb is the same.

 

  RSD Below Target.
Nothing Rejected.
    RSD Above Target.
Third Rep Rejected
    RSD Above Target. Third and Fifth Reps Rejected.
         
  Rep. TOC (ppb)     Rep. TOC (ppb)     Rep. TOC (ppb)
  1 1533     1 1533     1 1533
  2 1625     2 1625     2 1625
  3 1555     3 1799     3 1799
  4 1566     4 1566     4 1566
  5 1577     5 1577     5 1900
  Average 1571.2     Average 1620     Average 1684.6
  Std Dev 34.19     Std Dev 105.36     Std Dev 158.18
  RSD 2.18%     RSD 6.50%     RSD 9.39%
Stats with Rejection Average N/A   Stats with Rejection Average 1575.25   Stats with Rejection Average 1574.67
Std Dev N/A   Std Dev 38.07   Std Dev 46.61
RSD N/A   RSD 2.42%   RSD 2.96%

 