How can calibration points be removed from a Lachat Omnion calibration?

Dokument-ID

Dokument-ID TE8486

Veröffentlichungsdatum

Veröffentlichungsdatum 23.10.2019
Frage
How can calibration points be removed from a Lachat Omnion calibration?
Zusammenfassung
Removing points from Lachat calibration
Antwort
The Calibration Results can be viewed by selecting the Calibration Results button from the left-hand side of the channel data display.


Within the Calibration Results window the Unused column can be checked for one or more calibration points.  This will remove them from the calibration. The calibration statistics will change as the points are removed.
 

After closing the Calibration Results window, the modified calibration can be applied to the run. Right-clicking within the channel data display (the window peaks are visible in) will reveal a drop-down menu. Choosing Rerun Peak Detection will reprocess the samples and standards with the modified calibration.


 