What is the proper procedure to delete and re-install the Flow Module on the sc200?

Dokument-ID

Dokument-ID TE8621

Veröffentlichungsdatum

Veröffentlichungsdatum 05.01.2022
Frage
What is the proper procedure to delete and re-install the Flow Module on the sc200?
Zusammenfassung
Proper procedure to delete and re-install the Flow Module on the sc200
Antwort

Remove power from the sc200. Remove the Flow Module from the sc200.
Power the meter back on with the sensor and module disconnected.
The meter will notify you that the device is missing. Press enter to acknowledge the message. The meter will display a selection screen with the missing device listed.
Press the Left Arrow Key to select the device. Press Enter.
The meter will delete the device from the meter and display a screen saying "No Devices Connected".
Power off the meter and reconnect the Flow Module and sensor. Power the meter back on.
The meter will re-install the Flow Module and Sensor. The meter should detect the module and you should hear an audible clicking from the sensor. .