What are the different Error Messages on the Flo-Mate 2000?

Dokument-ID

Dokument-ID TE8659

Veröffentlichungsdatum

Veröffentlichungsdatum 20.12.2021
Frage
What are the different Error Messages on the Flo-Mate 2000?
Zusammenfassung
Error Messages on the Flo-Mate 2000
Antwort

Error Messages:

Low Bat:
Indicates low batteries (Replacement Guide)
Replace the batteries (Battery Life Guide)


Noise:
Indicates excessive electrical noise is present in the flow which will interfere with normal operation. This will cause the display to blank out.
Note:The noise flag usually comes on for few a seconds after the sensor is submerged even though there is no noise present. This is normal.




Con Lost:
Indicates that either the sensor electrodes are out of the water or they have become coated with oil or grease. After 5 minutes, the unit will turn itself OFF. If the electrodes are coated, clean them (Cleaning Guide).